Coriolis flow meter for carbon dioxide

Liquid phase and dosing

CO2 is stored as a liquid under pressure. Once pressure drops below equilibrium, part turns to gas and dry ice forms, so the meter goes upstream of the regulator.

Open configurator Request a quote
DN08-DN50up to 39,000 kg/h-20...+100 °Cup to 10 MPa±0.1-0.2%
Charging station: sensor in the cabinet
Charging station: sensor in the cabinet

What happens with CO2

Liquid CO2 at 20 °C exists above 1.5-2.0 MPa depending on purity and temperature. Density is 770-1000 kg/m³. On expansion part of the liquid boils, temperature falls and dry ice may form.

Dry ice clogs the sensor and valves. The meter therefore goes where pressure is above equilibrium, and the regulator and safety valves are downstream.

For beverage filling and carbonation the mass of CO2 matters, not volume: density in the cylinder and in the line differ.

Typical points and sizes

Typical points and sizes
PointFlow, kg/hDNExecution
Carbonation100-3,000DN08-DN25T
Cylinder filling1,000-10,000DN25-DN40T
Truck loading5,000-30,000DN40-DN50T
Reactor dosing50-1,000DN04-DN15T
100 - 10 000 кг/чDN01DN08DN15DN25DN50DN80DN150DN250
Meter range by DN and typical band for liquid CO2. 100-10 000 кг/ч, логарифмическая шкала.

What to check before ordering

  1. Inlet pressure above equilibrium at working temperature.
  2. No dead legs or pockets where dry ice can start.
  3. Line temperature and blockage risk.
  4. Drain and purge before service.
  5. Seal and tube materials for food-grade CO2.

Liquid or gas

In the liquid phase the meter measures mass and density steadily. If the flow after the regulator is two-phase, readings will drift: this is physics, not a meter fault.

For food-grade CO2, Tri-Clamp connections and the straight-tube sensor L are available; they are easier to clean and purge.

When Coriolis is not needed

If only total gas volume at the outlet is needed, a gas meter is simpler. Coriolis is used in the liquid phase and where accurate mass flow matters.

Frequent questions

Why upstream of the regulator?

Downstream pressure drops and CO2 flashes. In two-phase flow mass is calculated with an error.

Can gaseous CO2 be measured?

Yes, gas accuracy is ±0.5-1.0%. Process lines usually need the liquid phase.

What about dry ice?

It forms on expansion and clogs the sensor. Back pressure, heating and no dead legs help.

Which connections?

Tri-Clamp and straight-tube L for food lines; flanges and threads for industrial lines.